ImPulse
Elite Patient Manual
Problem Possible Cause Solution
No battery in unit, or
battery requires replace-
ment.
Battery not inserted
correctly.
Battery may be dis-
charged.
Brass/CGA sealing washer
is missing from unit.
Worn brass/CGA sealing
washer.
Unit not properly secured
to cylinder, or securing
knob not tight.
PULSE FLOW/CONTINUOUS
FLOW switch in CONTINU-
OUS FLOW position.
Cannula tubing is kinked or
not properly adjusted.
ON/OFF power switch set
to OFF position.
Oxygen cylinder is empty.
Oxygen cylinder valve is
closed.
If your unit fails to operate as described in this manual, first review the following
chart for possible causes and solutions. If necessary, contact your Equipment
Provider.
Troubleshooting Guide
Power switch ON
but no lights
display.
Intermittent power,
intermittent
pulsing, lights flash,
alarm sounds.
Oxygen leaks from
unit.
Constant flow of
oxygen from unit.
Unit does not
deliver oxygen.
Any other unit
problem.
Replace battery.
Insert battery properly by
referencing polarity guide on
inside of battery compartment
door.
Replace Alkaline battery with
new D-cell Energizer battery.
Ensure that the brass/CGA
sealing washer is present at the
oxygen cylinder connection on
the unit.
Replace brass/CGA sealing
washer.
Close cylinder valve and verify
that the unit alignment pins are
correctly fitted to the cylinder
alignment holes and that the
securing knob is tight.
Position switch to PULSE FLOW
position.
Check or replace tubing as
necessary.
Set ON/OFF power switch to ON
position.
Replace cylinder.
Open cylinder valve a minimum
of one turn counterclockwise.
Contact your Equipment
Provider.