ImPulse
Elite Patient Manual
Turning Off the Unit
1. Turn the oxygen cylinder valve fully clockwise to stop oxygen flow.
2. Set the ON/OFF power switch on the ImPulse Elite unit to the OFF
position.
Note
If you do not set the unit’s ON/OFF power switch to OFF when the unit is not in use, the
alarm sounds until the battery fully discharges.
Alarm/Light Indicators
Setting
Light Alarm
Indicates
OFF
No No Unit is not on, battery life is preserved.
ON +
(Flow
Selection)
Green No Pulse flow.
Yellow &
Green
Yes
(Intermittent)
Battery almost discharged. Prepare to replace
battery within 20-30 minutes.
Red
Yes
(Continuous)
No inspiration detected for approximately 30
seconds (cannula may be disconnected, tubing
kinked, or unit remains on when not in use).
BAT.
TEST
Green No Sufficient battery life.
Yellow &
Green
No
Battery almost discharged. Prepare to replace
battery within 20–30 minutes.
No No No battery or discharged battery in unit.
Alarm Light Indicators
The following chart provides an explanation of what each alarm light
indicates:
12